Model |
Gentier 48E |
Gentier 48R |
Fluorescence Channels |
4 |
2 |
Sample Throughput |
48 |
Authoritative Certifications: |
CE, FDA |
Compatible Fluorophores |
Channel 1: FAM, SYBR Green I, SYTO 9, EvaGreen, LC Green. Channel 2: HEX, VIC, TET, JOE. Channel 3: ROX, Texas Red. Channel 4: Cy5. |
Channel 1: FAM, SYBR Green I, SYTO 9, EvaGreen, LC Green. Channel 2: HEX, VIC, TET, JOE.
|
Lightsource |
High-brightness, long-life, maintenance-free LED light source |
Detector |
Photodiodes (PDs) |
Method |
Simultaneous well-by-well scanning of all fluorescent channels, without edge effects |
Detection Duration |
Detection of 48 wells of all channels completed in 2s. |
Heating & Cooling Rates |
Maximum heating rate of 8.0℃/s; Maximum cooling rate of 6.2℃/s. |
Temperature Accuracy |
≤0.1℃ |
Temperature-Controlled Programs |
Supporting conventional PCR, gradient PCR, touchdown PCR, long-range PCR, etc. |
Analytical Functions |
Qualitative analysis, absolute quantitative analysis, relative quantitative analysis, endpoint fluorescence analysis, dissociation curve analysis, and SNP analysis, etc. |
Sample Linearity & Repeatability |
Linearity: /r/≥0.999
Repeatability: cycle threshold (Ct) value CV≤0.5% |
Control method |
Stand-alone operation: 7-inch touchscreen control; Network operation: PC software control via direct connection or LAN (lcal area network) |
Suitable Consumables |
0.2mL transparent skirtless 48-well plate, 8-tube strips, and single tubes |
Power Failure Protection |
Automatic recovery of the experiment and other functions when the power is on again after cutting off, without waiting for the power-on of the computer or software control. |
Data storage |
> 1000 experimental data for stand-alone operation |
Customized Reporting |
Lab report templates of various industries pre-stored Open and universal reporting function, user-definable report content and form |
LIS Functions |
Open data port, synchronized interconnection with LIS system |
Stand-alone Analysis Results |
No need for an extra computer: immediate analysis of experimental data on the instrument; direct printing function (optional) to output the sample amplification curve and CT value, simply by connecting to the special thermal printer. |
Size and Weight |
260mm*400mm*260mm(W*D*H),11 kg |
Power supply |
AC 220 V (±10%), 50 Hz, 600 VA. |